Skip to main navigation Skip to search Skip to main content

Critical failure of a percutaneous discectomy probe requiring surgical removal during disc decompression

  • Richard Domsky
  • , Michael E. Goldberg
  • , Robert A. Hirsh
  • , Denise Scaringe
  • , Marc C. Torjman

Research output: Contribution to journalArticlepeer-review

Abstract

Objective: We report a complication while performing a percutaneous disc decompression at the L4-L5 level using a Dekompressor Percutaneous Discectomy Probe. Case Report: A 54-year-old male was referred to the pain clinic for possible percutaneous disc decompression. For the procedure the Dekompressor unit was inserted over the stylette and a percutaneous disc decompression was performed for 1 to 2 minutes using the channeling technique. Approximately 3/4 mL of disc nucleus was successfully removed. Upon withdrawal of the Dekompressor unit, it was noted that the probe was no longer connected to the device handle. Fluoroscopic imaging showed that approximately 4 inches of the probe remained in the patient and that the auger's distal end was still lodged in the disc. An incision was made, and the auger was successfully removed by a neurosurgeon. Conclusions: The patient had an uneventful recovery. Manipulation of the auger should be performed in a linear motion as best as possible and under fluoroscopic guidance.

Original languageEnglish (US)
Pages (from-to)177-179
Number of pages3
JournalRegional Anesthesia and Pain Medicine
Volume31
Issue number2
DOIs
StatePublished - Mar 2006
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Anesthesiology and Pain Medicine

Fingerprint

Dive into the research topics of 'Critical failure of a percutaneous discectomy probe requiring surgical removal during disc decompression'. Together they form a unique fingerprint.

Cite this